Diagnostic X-ray Unit
USER MANUAL
31
Exposure Data Table
[CHART FOR EQUINE]
- 26 Inch (66cm) SID
- 400 Speed Rare Earth Film / Screen
Note:
1. This exposure data table can considerable different
accordingly development condition.
2. This exposure data table can different kV and mAs
accordingly a object body situation.
ANATOMY
VIEW
High Contrast
ESD/Radiograph
(mGy)
Mid Contrast
ESD/Radiograph
(mGy)
kV
mAs
kV
mAs
NAVICULAR
AP
74
1.6
0.10
78
1.25
0.08
LAT
74
1.25
0.08
78
1.0
0.07
P3
74
1.0
0.06
-
-
-
FETLOCK
AP
74
1.6
0.10
78
1.25
0.08
LAT
70
1.25
0.08
78
1.0
0.07
KNEE
AP/FLEX
74
1.6
0.10
78
1.25
0.08
LAT/OBI
74
1.25
0.08
78
1.0
0.07
SPLINT BONE
LAT
70
1.0
0.05
76
0.8
0.05
HOCK
AP/HIGH
74
2.0
0.13
78
1.6
0.11
AP/LOW
74
1.6
0.10
78
1.25
0.08
LAT
74
1.25
0.08
78
1.0
0.07
STIFLE
LAT
-
-
-
80
3.2
0.24
PA
-
-
-
80
5.0
0.37
ELBOW
AP
-
-
-
80
3.2
0.24
OTHER
-
-
-
80
2.0
0.15
Summary of Contents for ANYPLX 3-BT
Page 4: ...Diagnostic X ray Unit USER MANUAL 4 Section 1 General Information ...
Page 22: ...Diagnostic X ray Unit USER MANUAL 22 n Rating Chart ...
Page 23: ...Diagnostic X ray Unit USER MANUAL 23 ...
Page 24: ...Diagnostic X ray Unit USER MANUAL 24 Section 2 Operation ...
Page 33: ...Diagnostic X ray Unit USER MANUAL 33 Section 3 Service Manual ...
Page 45: ...Diagnostic X ray Unit USER MANUAL 45 Section 4 Schematics ...
Page 46: ...Diagnostic X ray Unit USER MANUAL 46 Block Diagram ...
Page 47: ...Diagnostic X ray Unit USER MANUAL 47 Power Driver Board ...
Page 48: ...Diagnostic X ray Unit USER MANUAL 48 Main Board CPU ...
Page 49: ...Diagnostic X ray Unit USER MANUAL 49 Main Board EP_IP FB ...