Implementation of Radiographs
1) Please wear a lead apron while performing an x-ray
2) Turn on the main switch.
3) If you need to warm up this unit, expose 3 times each following the 5)
chart in page.27.
4) Select the technical mode : Manual or APR mode.
- Choose the most appropriate techniques for the exposure setting value
both the kV by means of the key and of the mAs by means of the key.
- When you select APR mode, and you want to store setting kV and mAs,
you should press the Store key for APR , and an audible buzzer sounds
at end of the store.
5) Exposure
① As making an exposure, press the hand switch half way down.
② During 1.5 seconds after this first press, the filament is heated and
green lamp (READY) lights.
③ Now depress the hand switch fully down and the exposure is given
immediately. The duration of the exposure is signaled by the orange
lamp (X-RAY) and an audible X-ray buzzer sound at the exposure.
Note:
1. The exposure must be made pressing the exposure switch
in two different times as above illustrated. However, it is
possible to press the switch all the way down at once
and the device will make the execution of the exposure
automatically after the heating (1.5 seconds) delay of the
filament heating.
2. The button should be depressed and held until the
exposure is fully completed.
3. In case some trouble occurs, the kV display indicates the
error code.
Summary of Contents for ANYPLX 3-BT
Page 4: ...Diagnostic X ray Unit USER MANUAL 4 Section 1 General Information ...
Page 22: ...Diagnostic X ray Unit USER MANUAL 22 n Rating Chart ...
Page 23: ...Diagnostic X ray Unit USER MANUAL 23 ...
Page 24: ...Diagnostic X ray Unit USER MANUAL 24 Section 2 Operation ...
Page 33: ...Diagnostic X ray Unit USER MANUAL 33 Section 3 Service Manual ...
Page 45: ...Diagnostic X ray Unit USER MANUAL 45 Section 4 Schematics ...
Page 46: ...Diagnostic X ray Unit USER MANUAL 46 Block Diagram ...
Page 47: ...Diagnostic X ray Unit USER MANUAL 47 Power Driver Board ...
Page 48: ...Diagnostic X ray Unit USER MANUAL 48 Main Board CPU ...
Page 49: ...Diagnostic X ray Unit USER MANUAL 49 Main Board EP_IP FB ...