Diagnostic X-ray Unit
USER MANUAL
43
5. It can be checked of heat sink equipped with LED lamp on the bottom side of
laser module after separating the bottom case from the equipment. Separate
the power connector of LED lamp from PCB.
6. After separating the power connector of LED lamp from PCB, draw a line by a
bright-color or white pen on the collimator side located just above the
right-end side of the existing LED lamp heat sink like the below photo. The
drawn line indicates the position of the heat sink of a install-planned new LED
lamp. In case of not making the line, it is able to cause the deviation between
the light of LED lamp and X-ray field. When installing the Heat Sink equipped
with a new LED lamp, assemble the heat sink as matching the right side of the
heat sink with the drawn line.
LED Lamp
Power
Connector
LED Lamp
Heat Sink
Position Line