Error Code and Corrective Actions
Code
Meanings of Error
Corrective Action
Err01
The ready contactor of the hand
switch is closed even it is not
pushed.
Reset main switch. If proper operation is
unavailable, change the hand switch.
Err02
It is displayed when the EP FB
signal is not detected in the state
of standby.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD.
Err03
It is displayed when the IP FB
signal is not detected in the state
of standby.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD.
Err04
It is displayed when the FIL FB
signal is not detected in the state
of standby.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD.
Err05
It is displayed when the Over
Current Protect is detected,
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD or replace
PWR BOARD.
Err06
It is displayed when the IP FB and
EP FB signal is high in the state of
exposure.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD.
Err07
It is displayed when the mAs error.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD or replace
PWR BOARD.
Err08
It is displayed when the mAs is
over in the state of exposure.
Reset main switch. If proper operation is
unavailable, check the IP FB level and adjust
it. Replace MAIN BOARD.
Err09
It is displayed when the Memory
error.
Reset main switch. If proper operation is
unavailable, replace MAIN BOARD.
Err10
It is display when overheated
(55 ℃) of HV Tank.
When the temperature reach 50 ℃, the
“CHECK” LED Lamp turns off then reset the
main switch.
Err11
It is display when thermistor circuit
open of HV Tank.
Repair HV Tank.
Summary of Contents for ANYPLX 3-BT
Page 4: ...Diagnostic X ray Unit USER MANUAL 4 Section 1 General Information ...
Page 22: ...Diagnostic X ray Unit USER MANUAL 22 n Rating Chart ...
Page 23: ...Diagnostic X ray Unit USER MANUAL 23 ...
Page 24: ...Diagnostic X ray Unit USER MANUAL 24 Section 2 Operation ...
Page 33: ...Diagnostic X ray Unit USER MANUAL 33 Section 3 Service Manual ...
Page 45: ...Diagnostic X ray Unit USER MANUAL 45 Section 4 Schematics ...
Page 46: ...Diagnostic X ray Unit USER MANUAL 46 Block Diagram ...
Page 47: ...Diagnostic X ray Unit USER MANUAL 47 Power Driver Board ...
Page 48: ...Diagnostic X ray Unit USER MANUAL 48 Main Board CPU ...
Page 49: ...Diagnostic X ray Unit USER MANUAL 49 Main Board EP_IP FB ...