Diagnostic X-ray Unit
USER MANUAL
42
Collimator LED Lamp Exchange Procedure
In case of No-light of collimator's LED lamp by a damage, please replace
according to the followings.
1. Turn "OFF“ the equipment
2. Place the collimator knob like the following photo, loosen the four bolts fixed
with collimator knob by a regular screwdriver as indicated by "D".
3. Loosen the eight bolts fixed on the both left and right of the upper and
bottom cases by a regular screwdriver.
4. Separate the both upper and bottom cases from the equipment. In case of
separating the upper case, gather collimator knob to the middle like the below
photo. And Separate the case by lifting only the right side in condition of
laying the case towards the left. Never lift the upper case by holding the
upside or downside of the case, which causes damaging collimator knob
resulting in deviating LED lamp from X-ray field.
D