User manual SEFELEC 5x Series
106
v1.06
If the current flowing in the element tested exceeds the cut-
out criteria set (in the ΔI or IMAX
mode), the device indicates that there a cut-out with cut-out of the high voltage (at the zero-
crossing of the sine wave), the display then indicates:
The coloured function key indicates the setting that caused the FAIL sanction of the measure-
ment.
WARNING
Before disconnecting the measured component, press the Stop button to
confirm the stoppage of the high voltage generation and discharge the
component: the device then performs a check of the residual voltage up
to a value of 30 VAC or DC
This action returns the user to the initial measurement screen either to change the settings or to
perform a new measurement by pressing the Start button.
STOP