THE FOLLOWING ERROR MESSAGES:
"SAFETY LOOP OPEN"
: the safety loop is not closed, there is no connection between
the points 1-9 or 2-10 of connection C5 on the rear panel of the device. Press OK to
acknowledge the message
"OVER RANGE"
: the continuity resistance of the sample under test is higher than
the measurement possibilities of the device (> 1,000 m
Ω)
"CONTINUITY ERROR
": The wires are not connected correctly, or the element
tested between these wires has a resistance that is too high (approximately a few
ohms) for the device to measure it.
IMPORTANT
The device is equipped with a thermal protection system. In case of extended
use of the High-current generator, the protection system will be activated.
In this case the "
OVER HEATED
" message will be displayed permanently on
the LCD screen.
Switch off the device and let it cool down for 10 minutes prior to resuming use.