User manual SEFELEC 5x Series
25
v1.06
Insulation resistance
(SEFELEC 1000-M; SEFELEC 56/506-D/S)
Measurement specifications
Programmable volt per volt from
20 to 1,000 VDC
Generation accuracy
±(1% + 1 V)
(+) pole of the generator
to ground
Dynamic stability for mains ΔV = ±10%:
Variation of the
voltage measured
< ±1%.
Maximum current in the measurement circuit:
2 mA - 20% +0%
Maximum capacity of the sample measured
< 100 µF (discharge time < 10 sec.)
Discharge resistance
2.2 kΩ
Resistance measurement range
Measurement scope
(Generator Utest/Umax) x 200
GΩ
(or 2
TΩ)
Standard version
Voltage
Measurement scope
100 V
100
kΩ to 20
GΩ
250 V
250
kΩ to 50
GΩ
500 V
500
kΩ to 100
GΩ
1,000 V
1
MΩ to 200
GΩ
Optional 2T
Ω version
Voltage
Measurement scope
100 V
100
kΩ to 200
GΩ
250 V
250
kΩ to 500
GΩ
500 V
500
kΩ to 1
TΩ
1,000 V
1
MΩ to 2
TΩ
Generation accuracy
Display with indication of the units kΩ, MΩ, GΩ and TΩ
2,000 points
Accuracy (in % of the reading, 1U = 1 point of display):
200 GΩ, basic version
±(1.5% + 1U)
Option -
2 TΩ and U
<= 200 Vdc
±(2% + 1U)
Option -
2 TΩ and U > 200
Vdc
±(1% x U test / 100 + 1U)
Capacity mode
1.00
MΩ at 200
GΩ
Accuracy Precision mode NORMAL ± 100
kΩ
Input impedance
10 MΩ ±1%