During measurement, it is recommended not to move or break the electrical contact with the
parts tested so as, on one hand, not to distort the test conditions (e.g.: duration of application)
and, on the other hand, not to affect the contact area by arcing due to the strong current.
WARNING
The passage of a strong current in an element with low resistance
can lead to a significant temperature rise. Do not touch the item test-
ed or the metal parts of accessories during or after the tests.
It is recommended to check the proper functioning of the device regularly by
connecting an element of known value to the measurement output and perform-
ing a measurement. In the absence of an element of known value, it is possible
to make a measurement on an open circuit and on a short-circuit to confirm the
consistency of the indications of the device.
After having connected the component for measurement applying the precautions described
above and selected the measurement settings required, return to the measurement screen: