OHM (OHMmeter) :
Sets the main thresholds unit (Ohm) and the ground continuity display function.
Caution : The action on this command will erase all previous recorded
thresholds (Reset to zero).
VOLT (VOLTmeter) :
Sets the main thresholds unit (Volt) and the ground continuity display function.
Caution : The action on this command will erase all previous recorded
thresholds (Reset to zero).
HTIM (HTIMe) <NR2> :
Sets a new value in second to the current hold time parameter to the parameter set
of the running function.
For instance : HTIM 3.2
RTIM (RTIMe) <NR2> :
Sets a new value in second to the current rise time parameter to the parameter set
of the running function
For instance : RTIM 10.5
FTIM (FTIMe) <NR2> :
Sets a new value to the current fall time parameter to the parameter set of the run-
ning function.
For instance : FTIM 0.5
HLIM (HLIMit) <NR3> :
Sets a new value to the high limit parameter to the parameter set of the running
function.
For instance :
MEG:HLIM 2.0E+6 (new value in Ohms)
HIP:HLIM 1.45E-4 (new value in Amps)
GND:HLIM 1.00E-1 (new value in milliohm)
The unity is automatic, given in Volt, Ohm or Amp and function dependent.
CB HLIM <NR3> :
Same command as HLIM, but reset the low limit parameter before to change the
high limit value.
LLIM (LLIMit) <NR3> :
Sets a new value to the low limit parameter to the parameter set of the running
function.
For instance :
MEG:LLIM 1.0E+6 (new value in Ohms)
HIP:LLIM 3.50E-6 (new value in Amps)
GND:LLIM 5.02E-2 (new value in milliohm)
The unity is automatic, given in Volt, Ohm or Amp and function dependent.
TIM (TIMe) AUT/PERM
Select the temporization mode. AUT for automatic, PERM for Permanent test until
a fault occurs.