Eagle EI-4000 Operating Manual Download Page 4

 

 

POUND (lb) AND KILOGRAM (kg) UNITS OPERATION 

The EI-4000 meter may operate using lb or kg units. DIP switch 
S1-3 selects the desired unit. If one wishes to switch to the other 
unit, simply flip the switch to the opposite setting; no recalibration 
is necessary. It is necessary to select the proper S1-3 switch 
setting so the unit selection works properly (i.e., do not calibrate the scale in kg if the S1-3 
switch is set to lbs). Also, all parameters that were set using the previous units are converted 
to the equivalent value of the newly selected unit. 
 

OPERATING MODES

1:GR   136.0 kg 
SEL SCL      MNU 

 

The EI-4000 meter has five operating modes to simplify user interactions with the meter. 
Configuration parameter P001 selects which operating mode, 1 thru 5, is enabled for all scale 
channels. 
 

MODE 1: FULLY MANUAL MODE 

Typical application: drum scales, bin and hopper scales - Remaining, tare, & gross weights 
are displayable. 
 

1. 

An empty scale should read zero. If it does not, press the ZRO (ZERO) 
pushbutton. If the ZRO pushbutton is not displayed, access the calibration 
mode to zero the scale. 

2. 

Place container on the scale. 

3. 

Press the SELECT pushbutton to go to the tare weight. 

4. 

If necessary, adjust the tare weight value via the arrow pushbuttons. 

5. 

Press the SELECT pushbutton to go to the remaining weight. 

 

 
MODE 2: ADJUSTABLE REMAINING 

Typical application: drum scales - Remaining & gross weights are displayable. 
 

1. 

An empty scale should read zero. If it does not, press the ZRO (ZERO) 
pushbutton. If the ZRO pushbutton is not displayed, access the calibration 
mode to zero the scale. 

2. 

Place container with product on the scale. 

3. 

Instrument will show the REMAINING weight and will flash the REM text. 

4. 

Use the UP and DOWN arrow pushbuttons to adjust the remaining weight 
displayed. 

5. 

Press the ENT pushbutton after making any adjustments; the user will no longer 
be able to adjust the remaining weight.  

 
If you need to adjust the REMAINING weight again, press and hold PB3 for 3 seconds, the 
REM text will start to flash and the remaining weight will now be adjustable. Press the ENT 
pushbutton when finished. 

 
 

Summary of Contents for EI-4000

Page 1: ...L EI 4000 DIGITAL WEIGH INDICATOR TRANSMITTER THIS MANUAL APPLIES TO EI 4000 FIRMWARE VERSION 6 31 ABOVE Revised 2015 04 07 EAGLE MICROSYSTEMS INC 366 CIRCLE OF PROGRESS POTTSTOWN PA 19464 EI 4000 MAN...

Page 2: ...uts two additional relay contacts and two 4 20mA analog outputs EI 2000 IO This board mounts on the EI 2000S CPU board It furnishes eight additional relay contacts and four remote switch inputs Note I...

Page 3: ...display at the top of this page Tare weight This is the weight of the container e g cylinder tank or barrel Remaining weight a k a net weight This is the weight of the product e g Chlorine that the c...

Page 4: ...If it does not press the ZRO ZERO pushbutton If the ZRO pushbutton is not displayed access the calibration mode to zero the scale 2 Place container on the scale 3 Press the SELECT pushbutton to go to...

Page 5: ...RE Typical application partially used chlorine cylinders Remaining tare gross weights are displayable 1 An empty scale should read zero If it does not press the ZRO ZERO pushbutton If the ZRO pushbutt...

Page 6: ...DOWN pushbuttons to adjust the gross scale calibration When satisfied that the weight is correct press the ENT pushbutton to save the new gross weight value EXTENDED MAIN SCREEN DISPLAYS rate daily us...

Page 7: ...eight display screen VIEWING ALL WEIGHT VALUES ON ONE SCREEN MAIN MENU SET LOG ALL ESC SETUP MENU CFG CAL RTC ESC G 0 0 R 150 0 R 75 0 G 0 0 If more than one channel is enabled a screen that displays...

Page 8: ...Low rate set point activated CHx DAILY HIGH High daily set point activated CHx DAILY LOW Low daily set point activated CHx SLOW REFILL Slow refill time has expired DAILY LOG The daily log displays the...

Page 9: ...he meter See the operating modes section of this manual for a detailed description of each one P010 lb kg display only This displays the engineering unit lb or kg selected by S1 3 The user can change...

Page 10: ...b kg 1 DCS302 300 140 0 5 0 2 1 2 60 27 2 150 136 2 EDS400 400 180 0 5 0 2 1 2 60 27 2 150 136 3 DS750 750 340 0 5 0 2 1 2 60 27 2 150 136 4 WT3600 1 ch 4000 1810 5 2 1 10 1000 454 2000 907 5 WT3600 2...

Page 11: ...l scale is 300lbs and auto zero is set to 0 5 the weight must be below 1 5lbs for the meter to automatically zero the scale P114 Push to Zero Auto zero PTZ AZ maximum range 2 5 10 20 50 or ALL When th...

Page 12: ...High level remaining weight set point The set points have a 1 hysteresis based on the full scale and an activation delay of three seconds For example if the full scale is 300lbs the set point is 10lbs...

Page 13: ...the load is inductive add a snubber RC circuit across the relay contact terminals to greatly extend the life of the relay contacts The relays on the EI 2000 CPU TB2 CH1 CH2 and EI 2000 EXP TB10 CH3 CH...

Page 14: ...ll below 4mA 0 lb kg or rise above 20mA until hardware limits are reached YES will limit the analog output to the 4mA to 20mA range SERIAL INTERFACE PARAMETERS P600 AOUT Gr NXT LAST ADJ ESC P610 A FS...

Page 15: ...ts the decimal point for the rate When the number cannot fit on the display the decimal point is shifted to accommodate the rate value and can be shifted to where no decimal point is displayed P910 R...

Page 16: ...t point The relay assigned can be shared with other set point relays Choices are OFF CH1 CH2 CH3 CH4 K1 through K8 K11 and K12 P933 D HIGH daily high level set point Entering a zero disables this set...

Page 17: ...s NO Scale inputs are not summed P991 SUM SEL sum selection RE Remaining weights are summed Gr Gross weights are summed P992 SUM EXT sum external EI 4000 YES Sum all active scales Up to eight scales a...

Page 18: ...le four is not enabled the CH4 pushbutton PB4 is reassigned as an escape ESC back to the calibration menu After a channel is selected for calibration the zero span calibration screen is displayed The...

Page 19: ...this must be taken into consideration when connecting the load to the analog outputs Refer to schematic above ANALOG 4mA and 20mA POINT ADJUSTMENT PARAMETERS Each analog output is calibrated to 4mA a...

Page 20: ...g weight e g 3mV V at 4000lbs must be known this is the scale FULL SCALE entry that s written on the EI routing sheet Scale calibration dead zero one for each active channel The scale mV V dead zero v...

Page 21: ...ENT SET NEW PASSWORD YES NO Power up functions are activated by first powering down the meter pressing the desired pushbutton combination and restoring power to the meter Restore all parameters and ca...

Page 22: ...acter 0aH The numbers shown in the text boxes in blue if printed in color are character indexes for programming purposes RS232 Wiring to PC DB9 Connector 00000000001111111111222222222233333333334444 4...

Page 23: ...RS 485 Wiring This is the wiring typical of an RS 485 interface...

Page 24: ...ere 250VAC 5x20mm time lag fuse Do not use any other rating VOLTAGE SELECT SWITCH S2 The voltage selection switch located on the EI 2000S CPU selects the 115V or 230V range EI 4000 METER SELECTION HEA...

Page 25: ...ading factory typical default This can be also performed by pressing the LOAD pushbutton in the scale full scale parameters P110 FACTORY DEFAULT Automatically loading default configuration calibration...

Page 26: ...ich is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale base which is 0 v...

Page 27: ...the scale base which is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale...

Page 28: ...INSTALLATION SPARE PARTS LCD1602 EI 110392 CPU P N 110394 EI 2000 IO 110396 EI 2000 EXP 110395 Interconnecting ribbon cable 8 111026...

Page 29: ...00645 WP1000 1 330 150 1 4 N RED GRN WHT BLK YEL 500736 or 500737 WP1000 2 661 5 300 1 4 N RED GRN WHT BLK YEL 500725 WP1000 3 1323 550 1 4 N RED GRN WHT BLK YEL 500722 DS750 750 340 2 1 N RED GRN WHT...

Page 30: ...ight 3 lb kg R float 2 0x101e 0x201e used weight 4 lb kg R float 2 0x1018 0x2018 rate 1 dwt dt R float 2 0x101a 0x201a rate 2 dwt dt R float 2 0x101c 0x201c rate 3 dwt dt R float 2 0x101e 0x201e rate...

Page 31: ...point 1 state 3 N A R uint16 1 0x1183 set point 1 state 4 N A R uint16 1 0x1184 set point 2 state 1 N A R uint16 1 Set point two state 1 ACTIVE and 0 INACTIVE 0x1185 set point 2 state 2 N A R uint16...

Page 32: ...Notes...

Page 33: ...INT MODE 1 P500 STOP BITS 1 OR 2 P703 SET POINT MODE 2 P501 PARITY EVEN ODD NONE P704 RTU NODE 1 32 P705 CHANNEL SETUP SCALE SEL P100 P200 P300 P400 FULL SCALE P110 P210 P310 P410 RESOLUTION P111 P211...

Reviews: