Eagle EI-4000 Operating Manual Download Page 12

 

P116  Threshold (0 to full-scale) 

This applies to operating modes 2 through 5. This sets the point 
(in lbs or kgs) that the meter will enable, depending on the 
operating mode, switching to remaining or enabling adjustment of 
the gross weight calibration or remaining weight.  
 
 

P117  Start (0 to full-scale) 

This parameter applies to operating modes 3 & 4. This is the 
value of the remaining weight (in lbs or kgs) when the scale 
channel is “started.” See the “operating modes” section of this 
manual. 
 
 

P2xx, P3xx & P4xx 

Parameters P2xx, P3xx and P4xx are identical to the P1xx set but control scale bases two 
through four, respectively. Inactive channel parameters are hidden. If the EI-2000-EXP is not 
installed, P3xx and P4xx parameters are not available. 
 

P500 Set point A mode selection 

 

Off 

Set points are disabled 

P116 THR    60.0 
NXT LAST ADJ ESC 

P117 START 150.0 
NXT LAST ADJ ESC 

P500 SP1mode  LO 

Lo(w)-level 

Low-level, remaining weight set point 

NXT LAST ADJ ESC 

Hi level 

High-level, remaining weight set-point 

 
The set points have a 1% hysteresis (based on the full-scale) and an activation delay of three 
seconds. For example, if the full-scale is 300lbs, the set point is 10lbs, and the low-level set 
point mode is selected, the relay is energized when the remaining weight is equal to or below 
10lbs for three seconds. The relay is de-energized when the remaining weight goes above 
13lbs. If the high-level set point is enabled, the relay is also energized at 10lbs or greater for 
3-seconds, but it’s de-energized at 7lbs. 
 

P501 Set point B mode selection 

This parameter is used if either the EI-2000-EXP or EI-2000-I/O 
board is installed or if the EI-1000 meter has only one channel 
enabled. The selections are identical to the P500 parameter. 
 

P502 SPevent 

Enables/disables event logging for set points one and two. 
 

P503 RLY 

This parameter is active when a EI-2000-IO board is installed.  
K1-K8:  

Set point relays for each channel are separate 

SHARE:  

Relays K1 & K2 on the EI-2000-IO board are 
assigned to SP1 and SP2, respectively. 

 

P501 SP2mode OFF 
NXT LAST ADJ ESC 

P502 SPevent YES 
NXT LAST ADJ ESC 

P503 RLY   K1-K8 
NXT LAST ADJ ESC 

Summary of Contents for EI-4000

Page 1: ...L EI 4000 DIGITAL WEIGH INDICATOR TRANSMITTER THIS MANUAL APPLIES TO EI 4000 FIRMWARE VERSION 6 31 ABOVE Revised 2015 04 07 EAGLE MICROSYSTEMS INC 366 CIRCLE OF PROGRESS POTTSTOWN PA 19464 EI 4000 MAN...

Page 2: ...uts two additional relay contacts and two 4 20mA analog outputs EI 2000 IO This board mounts on the EI 2000S CPU board It furnishes eight additional relay contacts and four remote switch inputs Note I...

Page 3: ...display at the top of this page Tare weight This is the weight of the container e g cylinder tank or barrel Remaining weight a k a net weight This is the weight of the product e g Chlorine that the c...

Page 4: ...If it does not press the ZRO ZERO pushbutton If the ZRO pushbutton is not displayed access the calibration mode to zero the scale 2 Place container on the scale 3 Press the SELECT pushbutton to go to...

Page 5: ...RE Typical application partially used chlorine cylinders Remaining tare gross weights are displayable 1 An empty scale should read zero If it does not press the ZRO ZERO pushbutton If the ZRO pushbutt...

Page 6: ...DOWN pushbuttons to adjust the gross scale calibration When satisfied that the weight is correct press the ENT pushbutton to save the new gross weight value EXTENDED MAIN SCREEN DISPLAYS rate daily us...

Page 7: ...eight display screen VIEWING ALL WEIGHT VALUES ON ONE SCREEN MAIN MENU SET LOG ALL ESC SETUP MENU CFG CAL RTC ESC G 0 0 R 150 0 R 75 0 G 0 0 If more than one channel is enabled a screen that displays...

Page 8: ...Low rate set point activated CHx DAILY HIGH High daily set point activated CHx DAILY LOW Low daily set point activated CHx SLOW REFILL Slow refill time has expired DAILY LOG The daily log displays the...

Page 9: ...he meter See the operating modes section of this manual for a detailed description of each one P010 lb kg display only This displays the engineering unit lb or kg selected by S1 3 The user can change...

Page 10: ...b kg 1 DCS302 300 140 0 5 0 2 1 2 60 27 2 150 136 2 EDS400 400 180 0 5 0 2 1 2 60 27 2 150 136 3 DS750 750 340 0 5 0 2 1 2 60 27 2 150 136 4 WT3600 1 ch 4000 1810 5 2 1 10 1000 454 2000 907 5 WT3600 2...

Page 11: ...l scale is 300lbs and auto zero is set to 0 5 the weight must be below 1 5lbs for the meter to automatically zero the scale P114 Push to Zero Auto zero PTZ AZ maximum range 2 5 10 20 50 or ALL When th...

Page 12: ...High level remaining weight set point The set points have a 1 hysteresis based on the full scale and an activation delay of three seconds For example if the full scale is 300lbs the set point is 10lbs...

Page 13: ...the load is inductive add a snubber RC circuit across the relay contact terminals to greatly extend the life of the relay contacts The relays on the EI 2000 CPU TB2 CH1 CH2 and EI 2000 EXP TB10 CH3 CH...

Page 14: ...ll below 4mA 0 lb kg or rise above 20mA until hardware limits are reached YES will limit the analog output to the 4mA to 20mA range SERIAL INTERFACE PARAMETERS P600 AOUT Gr NXT LAST ADJ ESC P610 A FS...

Page 15: ...ts the decimal point for the rate When the number cannot fit on the display the decimal point is shifted to accommodate the rate value and can be shifted to where no decimal point is displayed P910 R...

Page 16: ...t point The relay assigned can be shared with other set point relays Choices are OFF CH1 CH2 CH3 CH4 K1 through K8 K11 and K12 P933 D HIGH daily high level set point Entering a zero disables this set...

Page 17: ...s NO Scale inputs are not summed P991 SUM SEL sum selection RE Remaining weights are summed Gr Gross weights are summed P992 SUM EXT sum external EI 4000 YES Sum all active scales Up to eight scales a...

Page 18: ...le four is not enabled the CH4 pushbutton PB4 is reassigned as an escape ESC back to the calibration menu After a channel is selected for calibration the zero span calibration screen is displayed The...

Page 19: ...this must be taken into consideration when connecting the load to the analog outputs Refer to schematic above ANALOG 4mA and 20mA POINT ADJUSTMENT PARAMETERS Each analog output is calibrated to 4mA a...

Page 20: ...g weight e g 3mV V at 4000lbs must be known this is the scale FULL SCALE entry that s written on the EI routing sheet Scale calibration dead zero one for each active channel The scale mV V dead zero v...

Page 21: ...ENT SET NEW PASSWORD YES NO Power up functions are activated by first powering down the meter pressing the desired pushbutton combination and restoring power to the meter Restore all parameters and ca...

Page 22: ...acter 0aH The numbers shown in the text boxes in blue if printed in color are character indexes for programming purposes RS232 Wiring to PC DB9 Connector 00000000001111111111222222222233333333334444 4...

Page 23: ...RS 485 Wiring This is the wiring typical of an RS 485 interface...

Page 24: ...ere 250VAC 5x20mm time lag fuse Do not use any other rating VOLTAGE SELECT SWITCH S2 The voltage selection switch located on the EI 2000S CPU selects the 115V or 230V range EI 4000 METER SELECTION HEA...

Page 25: ...ading factory typical default This can be also performed by pressing the LOAD pushbutton in the scale full scale parameters P110 FACTORY DEFAULT Automatically loading default configuration calibration...

Page 26: ...ich is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale base which is 0 v...

Page 27: ...the scale base which is 5V 5 2 S Positive signal Positive scale signal input 3 S Negative signal Negative scale signal input 4 E Negative excitation Negative power supply voltage return for the scale...

Page 28: ...INSTALLATION SPARE PARTS LCD1602 EI 110392 CPU P N 110394 EI 2000 IO 110396 EI 2000 EXP 110395 Interconnecting ribbon cable 8 111026...

Page 29: ...00645 WP1000 1 330 150 1 4 N RED GRN WHT BLK YEL 500736 or 500737 WP1000 2 661 5 300 1 4 N RED GRN WHT BLK YEL 500725 WP1000 3 1323 550 1 4 N RED GRN WHT BLK YEL 500722 DS750 750 340 2 1 N RED GRN WHT...

Page 30: ...ight 3 lb kg R float 2 0x101e 0x201e used weight 4 lb kg R float 2 0x1018 0x2018 rate 1 dwt dt R float 2 0x101a 0x201a rate 2 dwt dt R float 2 0x101c 0x201c rate 3 dwt dt R float 2 0x101e 0x201e rate...

Page 31: ...point 1 state 3 N A R uint16 1 0x1183 set point 1 state 4 N A R uint16 1 0x1184 set point 2 state 1 N A R uint16 1 Set point two state 1 ACTIVE and 0 INACTIVE 0x1185 set point 2 state 2 N A R uint16...

Page 32: ...Notes...

Page 33: ...INT MODE 1 P500 STOP BITS 1 OR 2 P703 SET POINT MODE 2 P501 PARITY EVEN ODD NONE P704 RTU NODE 1 32 P705 CHANNEL SETUP SCALE SEL P100 P200 P300 P400 FULL SCALE P110 P210 P310 P410 RESOLUTION P111 P211...

Reviews: