P114 PTZ 2
NXT LAST ADJ ESC
P110 Full-scale (10 to 9900)
This sets the full-scale in the units selected by the lb/kg switch,
S1-3. This is used to limit set points and other parameters that
are entered using lb or kg weighing units. This parameter is
available only when the scale selection parameter, P100, is set to “user.”
P111 Resolution (0.0001 to 500)
This sets the weight division for the unit selected by the lb/kg
switch. The choices are 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, and 5.
P112 Average (1, 2, 4, 8, 12, 16, or 20)
This parameter sets the number of scale weight acquisitions
before the display is updated to the weight on the scale. Higher
numbers yield a more stable less responsive meter reading.
P113 Auto-zero (off, 0.01, 0.02, 0.03, 0.05, 0.1, 0.2, 0.3, 0.5, 1, 2)
Auto zero is used to automatically zero the meter (same as pressing the zero pushbutton).
This setting controls how much of scale range in percentage will automatically be zeroed
when the load (e.g., container, tank or cylinder) is removed from the scale. All negative
values are zeroed regardless of this setting unless it is set to off.
Example: If full-scale is 300lbs and auto-zero is set to 0.5 (%), the
weight must be below 1.5lbs for the meter to automatically zero
the scale.
P114 Push to Zero/Auto-zero PTZ/AZ maximum range (2, 5, 10, 20, 50, or ALL).
When the load is removed from the scale, in an ideal world, the weight returns to zero.
However, usually the meter reading is above zero. If the auto-zero is enabled (Px13) and the
weight is within its range, the meter will automatically zero the weight. If it’s not, the user
should press the ZRO (zero) pushbutton for the selected channel.
The PTZ/AZ parameter, set in percentage of full-scale, selects
the maximum amount of weight that is permitted to be zeroed
either automatically or by pressing the zero pushbutton.
Example: If full-scale is 300lbs and
PTZ/AZ
is set to 2 (%), the weight must be below 6 lbs for
the meter to allow the automatic zero (AZ), if enabled, to zero the scale or the ability to zero
the scale via the ZRO pushbutton. Note: If the threshold (Px16) weight is exceeded
(operating modes 2-5), the automatic zero and pushbutton zero are inhibited.
Note: The “ALL” setting allows unlimited zeroing ability, which is not a recommended setting
since the scale load-cell may be shifting too much.
P110 FS 300
NXT LAST ADJ ESC
P111 RES 0.5
NXT LAST ADJ ESC
P112 AVG 4
NXT LAST ADJ ESC
P113 AZ 1
NXT LAST ADJ ESC