ATC770 Microprocessor-Based Pressure/Process Controller
55
3. The display values for analog inputs are scaled from 0 to 25,000 counts; it is also linear for RTD
input.
4. Use the
▼
or
▲
keys to select a display value from 0 to 10 and to check the linearity of output
circuit at 0%, 10%, 90% and 100% of full scale value.
5. If the values do not correspond with the values in the Calibration Parameters Summary Table
below, use the
▼
or
▲
keys to correct the value displayed.
6. When all the appropriate values are correct, set the dip switches to operating mode.
CALIBRATION PARAMETERS SUMMARY
Parameter
Circuit
Input Type
Range
Value
Note
CAL ATC
(2)
PL.020
Primary Input
Current
Zero
0 mA
PH.020
Primary Input
Current
Full scale
20 Ma
P.020
Primary Input
Current
Verify
(3)
PL.05
Primary Input
Voltage 0-5V
Zero
0 V
PH.05
Primary Input
Voltage 0-5V
Full scale
5 V
P.05
Primary Input
Voltage 0-5V
Verify
(3)
PL.010
Primary Input
Voltage 0-10V
Zero
0 V
PH.010
Primary Input
Voltage 0-10V
Full scale
10 V
P.010
Primary Input
Voltage 0-10V
Verify
(3)
ML.CUR
Main analog output
Current
Zero
-5 mA
MH.CUR
Main analog output
Current
Full Scale
25 mA
M.CUR
Main analog output
Current
Verify
(4)
ML.VOL
Main analog output
Voltage
Zero
-12.5 V
MH.VOL
Main analog output
Voltage
Full scale
12.5V
M.VOL
Main analog output
Voltage
Verify
(4)
SL.CUR
Second analog output
Current
Zero
-5 mA
SH.CUR
Second analog output
Current
Full scale
25 mA
S.CUR
Second analog output
Current
Verify
(4)
SL.VOL
Second analog output
Voltage
Zero
-12.5 V
SH.VOL
Second analog output
Voltage
Full scale
12.5 V
S.VOL
Second analog output
Voltage
Verify
(4)
DEFLT
Load default calibration
DO NOT ATTEMPT TO MODIFY
and code
UNDER ANY CIRCUMSTANCES
7. If the values CANNOT be made to correspond with the values in the Calibration Parameters
Summary Table, the instrument must be sent to Dynisco for repair or re-calibration.
8.3
RS-485 (O
PTIONAL
)
The ATC770 is available with an RS485 Digital communications port. The configuration parameters
for this option are found in the Group 3 parameters only if this option is included. The ATC770,
Summary of Contents for ATC770
Page 19: ...ATC770 Microprocessor Based Pressure Process Controller 19 Fig 2 ATC770 Outline Drawing...
Page 23: ......
Page 30: ...30 Fig 6 Remote Set Point Jumper Location Board E...
Page 84: ...NOTES...
Page 85: ...NOTES...
Page 86: ...NOTES...
Page 88: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...