4.5 Customizing the Program
Following examples show how to modify the parameters, save and compile the program before making it
ready to send to the logger.
Procedure to Modify the Program:
Open CR Basic Editor by clicking on CRBasic icon
or your favorite text editor.
Open the file IMET_MP2_Main.CR2 in the editor.
Now identify the site-specific or timing loop variables at the top of the program as shown in the
previous above.
Modify values assigned to the constants as required.
Save the program under a different file name (ImetSiteA.cr2) so that the original program is not
modified.
Now the new program ImetSiteA.cr2 is ready to send to logger.
Example 1:
InteliMet Advantage 6 weather station is setup on top of a tower whose height is 10m. Roughness
parameter of the location of the installation is found to be 0.001m and the specific air heat capacity is
1002.0 J/KgC. Modify the program IMET_MP2_Main.CR2 and send to logger so that the weather station
can calculate ET using the Lascano-VanBavel RCM algorithm.
From the program it is clear that the required station parameters are,
zom = 10 m
zot = 0.001 m
lev = 2.44 x 10
6
has = 1002.0 J/KgC
Open the program in IMET_MP2_Main.CR2 in CRBasic or favorite text editor. As shown below are the
default parameters.