www.drawell.com.cn
2.2.5 instrument working condition and performance indicator
2.2.5.1 instrument working condition
Ambient temperature : 10
℃~
30
℃
Relative humidity : ≤ 70%
Power supply : AC220V±22V 250VA 50Hz±1Hz
2.2.5.2 instrument storage conditions
Ambient temperature : -20
℃~
+55
℃
Relative humidity : ≤80%
Note
: In order to assure the normal working and the precision of the experiment result,
please operate and store the instrument in the stipulated condition. Otherwise there may
be damage to the instrument and harm the operator’s safety and life.
2.2.5.3 instrument performance indicator
Model:
K960
-A
;
K960
-B
;
K960
-C
;
K960
-D
Sample size: 96*0.2mL(A);54*0.5mL(B);96*0.2mL+77*0.5mL(C);384well(D)
Temperature range: 0
℃—
99.9
℃(
Rt
≤
30
℃)
Temperature ascending speed:
≥
3
℃
/s
Temperature descending speed:
≥
2.8
℃
/s
Temperature uniformity:
≤±
0.4
℃
Temperature accuracy:
≤±
0.1
℃
Gradient temperature range: 30
℃—
99
℃
Gradient temperature width: 2
℃—
30
℃
heat lid temperature: 20
℃—
110
℃
Style of temperature control:block
;
calculated
Maximum number of program units: 200
Maximum number of cycles: 99
Graphical interfaces:5.7inch
,
320*240pels LCD
Communication interface: USB2.0
Exterior size: 380mm(L)*270mm(W)*250mm(H)
Net weight: 7.2kg