6. Question&Answering
6.1 Q: Is the “Season control mode” should be setting before programming?
A: Before programming, select the “Set” on the main menu interface to get into the
“Parameter setting” interface, select the corresponding mode of the current area
season on the “Season control mode” option.
NOTE: System will default the last operation season mode.
Click here to enter the "
Parameter Setting"
Season control mode
6.2 Q: Is the “grads” function the same with the “ascending” and the “descending”
function?
A: The grads function is presented in the middle row of the sample base. The
temperature presented on the right and left side is descending or ascending. You can
refer to the specific corresponding figure on the grads temperature table. This function
can significantly increase the efficiency of temperature controlling when the
experiment is still in the stage of exploring.
The ascending and descending of temperature means when another repetition is
made, the temperature will add or subtract a corresponding temperature based on the
preset temperature. For example, the starting temperature is 55
℃
. If the temperature
ascend (descend) 1
℃
, then the temperature in the second circle is 56
℃
(54
℃
for
descending). And in the third circle the temperature will be 57
℃
for the ascending and
53
℃
for the descending.