www.drawell.com.cn
4.2.2 problem reason analysis and corresponding solving method
No.
Phenomenon
Possible cause
Processing Procedure
1
No display and the beep
missed when power on
Disconnected power
supply
Check power supply for correct
connection
Bad fuse
Replace it(250V 2.5A Φ5×20)
Switch failure
Replace it
Other
Contact distributor or
manufacture
2
Display of Error! Flicker
and beep alarm
Possibly reasons
1. Open circuit in cooling
block sensor
2. Short circuit in cooling
block sensor
3. Open circuit in hotlid
4. Short circuit in hotlid
5. Open circuit in radiator
sensor
6. Short circuit in radiator
sensor
Contact distributor or
manufacture
3
Cooling speed slows
down obviously or
cannot cooling
Ventilation jam
Clear the jam
Cooling block failure
Contact distributor or
manufacture
Fan failed or stopped
4
Heating speed slows
down obviously or
cannot heating
Cooling block failure
5
Heating and cooling
terminated
Bad sensor
Contact distributor or
manufacture
Malfunction of all cooling
blocks
6
Lid cannot be heated
Lid heater failure
Contact distributor or
manufacture
Lid sensor failure
Bad contact between
connectors
7
Abnormal characters
displayed
Bad contact between chip
and its socked
Contact distributor or
manufacture
Chip malfunctioning
8
Inactive keys
Film panel malfunctioning
Contact distributor or
manufacture