Instrument Maintenance
To keep the instrument work in good condition, constant maintenance is
needed.
Daily Maintenance
1, Check the compartment
After measurement, the cuvettes with sample solutions should be
taken out of the compartment in time. Or the volatilization of the
solution would make the mirror go moldy. Users must pay more
attention to the corrosive samples and liquid easy to volatilize. Any
solution remains in the compartment should be wiped off immediately.
2, Surface Clean
The cover of the instrument is with paint. Please use wet towel to wipe
off the drips on the surface immediately. Organic solution is forbidden
to be used to clean the cover. Please wipe off the dirt on the cover
timely.
3, Clean the cuvettes
After every test or after a solution change, the cuvettes should be
cleaned carefully, or the remains on the surface would cause measuring
error.
Troubleshooting
1, No response after power on
Possible Cause
Solution
Bad contact in power supply
Improve the contact
Fuse melt
Replace a new fuse
2, No Stable Reading
Possible Cause
Solution
No enough pre-warm
Increase the pre-warm time
Glass cuvettes used in UV Range
Use Silicon Cuvettes in UV Range
unstable Sample
Improve the sample
Much higher sample concentration
Dilute the sample
Low voltage or unstable power supply
Improve the power condition
Lights defect
Replace a new lamp
Light used up
Replace a new lamp
3, Worse Repeatability
Possible Reason
Solution
Unstable sample
Improve the sample
Cuvettes polluted
Clean the cuvettes