41
Fig 97
4. Press
【
F3
】
to set tolerance (Fig 98). Input the value by pressing the
numeric keypad.
Fig 98
5. Press
【
ZERO
】
for blanking.
6. Put the sample (calibrated holmium liquid) into Main Light Path.
Press
【
START
】
to check. The results will be displayed on the
screen (Fig 99).
If the discrepancy between the results and the
calibrated values is not more than the tolerance, “pass” will be
displayed after the test results. Otherwise, “fail” will be displayed.
Fig 99
6.The result can be saved,loaded and printed by pressing
【
SAVE
】,【
LOAD
】
and 【
】