Step 4: Remove the cover of the Lamp Chamber
Unscrew the screws of the Lamp Chamber indicated in Fig.103 and remove
its cover. (Fig 103)
Fig 103
Step 5: Replace Lamps
Fig 104 Top View of Lamp Chamber
1) Replace D2 Lamp
Unscrew the 2 screws on the D2 Flange (Indicated in the Red Circles in
Fig 105), unplug the power connector
(
Indicated in the Red Square in
Fig.104
)
in the Power Board and remove the D2 lamp. Draw on the
Cotton Glove and replace a new lamp. Fix the 2 screws and plug the
connector again.
The D2 Lamp is pre-aligned, so there’s no need to re-adjust the position
of it. The Facula should focus on the center of the Slit.
D2 Lamp
Motor
W Lamp
D2 Lamp
Connector