39
Fig 91
3.
Press
【
F2
】
to select test mode (Abs or %T, Fig 92).
Fig 92
4.
Press
【
F3
】
to set tolerance (Fig 93).Input the value by pressing the
numeric keypad.
Fig 93
5.
Press
【
ZERO
】for
Blanking.
6.
Put the sample (calibrated neutral density filter) into Main Light Path.
Press
【
START
】
to check. The results will be displayed on the screen
(Fig 94).
If the discrepancy between the results and the calibrated
standards is not more than the tolerance, “pass” will be displayed after
the test result. Otherwise, “fail” will be displayed.
7.
The result can be saved,loaded and printed by pressing
【
SAVE
】 ,
【
LOAD
】and 【
】.