9
automatically. However the
【
START
】
must be pressed in other measurements
such as DNA/Protein, Muli WL and Quantitative etc.
B. Take measure in WL Scan
a.
After all scan parameters are entered, put the reference cuvette with
reference solution
into the Reference Light Path and the sample cuvette with
sample solution
into the sample light path,Press
【
START
】
to scan
.
b.
(Recommended) After all scan parameters are entered, put the
reference cuvette with
reference solution
into the Reference Light Path and the
sample cuvette with
reference solution
into the Sample Light Path, Press
【
ZERO
】
to obtain the user baseline
.
Then take out the sample cuvette,
replace the
reference solution
with
sample solution
after flushing the cuvette
completely. Put the sample cuvette into the Sample Light Path. Press
【
START
】
to scan.
※
Set wavelength
(Example: set wavelength in “Basic mode”)
Press
【
SET
λ】
(Fig 10).
Fig 10
Use numeric keypad to input wavelength (Fig 11).
Fig 11
Press
【
ENTER
】
to change the wavelength from 656.1nm to
450.0nm,and then blank; After blanking, the screen displays as Fig
12.