Section 2
Setup and Installation
Gas Analyzer
Exhaust
Because the monitor withdraws a portion of the patient breathing system
gas, sampling gas must be exhausted to the scavenging system.
If you have an open reservoir scavenging system, connect the exhaust line
leading from the rear of the monitor to the scavenger as shown in the
figure below:
If you have a scavenger interface for passive systems, connect the exhaust
FITTING
EXHAUST
PORT ON REAR
OF V3000
(REAR VIEW OF SCAVENGER)
EXHAUST
SU30008
line leading from the rear of the monitor to the scavenger using a
scavenger adapter, as shown in the figure below:
SCAVENGER
SCAVENGER
ADAPTER
SU30049
2-9