Section 3
Routine Maintenance
Overview
This section includes cleaning and replacement procedures and instructions
on the use of the Vitalert 3000 Series maintenance screens.
Cleaning and
Sterilization
Clean and sterilize the Vitalert 3000 Series monitor and its parts according
to the guidelines below. Follow your institution’s policies regarding specific
methods and agents for cleaning and sterilization. Determination of the
need and frequency of sterilization of any particular component is the
responsibility of the user institution. Sterilization procedures should be
performed according to procedures established by the user institution,
following the specific instructions provided by the manufacturer of the
sterilizing equipment or agent to be used. Such policies, procedures, and
instructions should ultimately be consistent with established principles of
clinical microbiology and infection control.
Surfaces
Clean painted, plated, and plastic surfaces of the Vitalert 3000 Series with a
soft cloth moistened with an aqueous germicidal cleaner. Mix the
germicidal cleanser in accordance with instructions provided by the
manufacturer. Do not use solvent cleaners or abrasive cleaning agents on
any surfaces of the monitor.
Do not use anesthetic agents for cleaning purposes.
CAUTION:
Do not allow liquid to enter the interior of the Vitalert 3000
Series monitor.
Pulse Oximeter
Sensor
Do not autoclave the sensor or interface cable. The sensor can be cleaned
by wiping the patient contact areas with a liquid disinfection agent or
alcohol. Do not use acetone or acetone/alcohol mixtures because they will
damage the sensor. Do not immerse the sensor or interface cable. Do not
allow liquid to enter the interior of the sensor.
Noninvasive
Blood Pressure
Cuff
Before cleaning or sterilizing the NIBP cuff, detach it from the extension
hose by twisting the Luer-lock fitting at the juncture of the cuff hose and
measurement hose. Clean the cuff with a liquid disinfection agent or mild
detergent solution.
Do not autoclave the cuff. It can be sterilized with ethylene oxide gas,
followed by at least 2 hours aeration in an appropriate aeration cabinet.
Do not autoclave the extension hose. It can be sterilized with ethylene
oxide gas (cold cycle), followed by at least 8 hours aeration in an
appropriate aeration cabinet.
3-1
Summary of Contents for Vitalert 3000 Series
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Page 4: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Page 6: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Page 16: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Page 36: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Page 38: ...RETURN TO CD ROM TABLE OF CONTENTS ...