Signature
Title/Dept.
Telephone ( )
Date
Ext.
COMMENTS:
Vitalert 3000 Routine Maintenance Checklist
State
Zip
Institution
Address
City
Model No.
Serial Number
Room No.
1.0 SPO2/PULSE
1.3 VERIFY O2SAT AND PULSE SIMULATOR
1.5 DURASENSOR READINGS
1.7 NO OXI PULSE WARNING
1.10 OXI PULSE LOW WARNING
1.13 OXI PULSE HIGH CAUTION
1.16 SPO2 LOW WARNING
1.19 SPO2 HIGH CAUTION
2.0 NIBP
2.3 BP CUFF LEAK ADVISORY
2.5 NIBP TEST PASSED
2.6 NIBP ACCURACY
2.10 NIBP SYS HI CAUTION
2.13 NIBP SYS LO CAUTION
2.15 STAT INTERVAL
Following completion of all calibrations/verifications, set the cursor to RESET HOURS
selection and press rotary dial to reset the Hours Run Since Last Service to zero.
After reseting hours to zero Vitalert 3000 shall be cycled through STBY.
NOTES:
Electrical Safety Testing ( to include chassis Leakage and Ground Continuity check )
shall be performed by qualified personnel on a quarterly basis.
3.0 CO2/SAMPLE FLOW
3.1.1 100 ml/min 25
3.1.2 200 ml/min 25
3.1.3 LINE BLOCK ADVISORY
3.2 CO2 (4.8% - 5.2%)
3.7 ETCO2 LOW CAUTION
3.11 ETCO2 HIGH CAUTION
3.14 APNEA - CO2 CAUTION
3.15 APNEA - CO2 WARNING
3.17 INSP CO2 HIGH ADVISORY