Section 3
Routine Maintenance
Maximum Flow
Adjustment
(not on all models or
earlier versions)
Maximum flow adjustment procedure:
1. Use the rotary dial to position the cursor on the
ADJUST FLOW
selection.
2. Press the rotary dial.
3. Rotate the dial until the flow meter connected to the exhaust port
reaches a maximum flow "plateau".
4. Adjust the variable flow restrictor next to the water trap reservoir until
the flow meter indicates 220 ml/min. Turn the screw to the left to
increase the flow or to the right to decrease it.
5. Press the rotary dial.
6. Proceed with the flow calibrations.
OXIMETER
SENSOR
BP CUFF
VARIABLE FLOW
RESTRICTOR
SU30048
3-9