5

Replacing the Speech Diaphragm or O-Ring

Draw the nose cup out of the groove in the 
screw ring.

Unscrew the screw ring with spanner 
(R 26 817).

Turn the socket wrench over and carefully 
insert it in the holes of the grille, taking care 
not to damage the foil on the diaphragm. 

Turn the speech diaphragm to release it and 
remove it. 

Remove the O-ring from the connector with 
sealing ring lifter (40 54 983). 

Check speech diaphragm and O-ring and replace if necessary.

When assembling the parts:

Press the edge of the nose cup into the groove in the screw ring and 
ensure that it engages all around. The edge of the nose cup must fit all 
around the full circumference of the screw ring. The center marks on the 
facepiece connector and on the nose cup must match.

Replacing the Exhalation Valve Disc

Remove the exhalation valve cover. 

Grip under the edge of the valve disc and pull the holding stud out of the 
hole.

Clean the valve seat. 

Moisten the new valve disc with water and press it into the hole until it 
engages. The valve disc must make even contact all around. 

Refit the exhalation valve cover so that it clicks into place.

Replacing the Inhalation Valve Disc

Pull the inhalation valve out by its tab. 

Remove the old valve disc and fit new disc. 

Press the inhalation valve into the mask connector until it makes even 
contact all around. The valve disc must not be pinched. The disc must 
face inwards, and the stem outwards.

Maintenance Intervals

Testing / Visual Inspection

The mask must regularly be serviced (see “Maintenance Intervals”).

Test Equipment and Test Accessories

— Test Kit (40 58 234) or Testor 2100 (R 53 400)
— Adapter (R 53 344) 
— Exhalation valve plug (R 53 349), or other commercially available test 

equipment.

Visual Inspection of the Speech Diaphragm

The speech diaphragm must not show any damages, e.g. deformations, 
holes, oxidations. Otherwise it must be replaced.
Please refer also to the table titled “Maintenance Intervals” on page 5. To 
replace the speech diaphragm, see “Replacing the Speech Diaphragm or 
O-Ring” on page 5.

Visual Inspection of the Exhalation Valve 

Remove the exhalation valve cover from the mask connector, and carefully 
inspect the valve disc. Valve disc and seat must be clean and undamaged. 
Clean or replace if necessary.

Work required

Bef

ore use

Af

ter use

Ever

y 6 mont

hs

Ever

y 2 year

s

Ever

y 4 year

s

Ever

y 6 year

s

Cleaning and disinfection

X

X

1

1 for un-used, stored mask in original packaging, otherwise every 6 months

Visually inspect, correct 
function and leak-check

X

X

1

Replace exhalation valve disc

X

Replace speech diaphragm

X

Visual check by wearer

X

WARNING!
Testing and maintenance is to be performed by an 
authorized Dräger service center or Dräger certified, 
Level II technician.

00721592_1.eps

Fig. 13

Leak Test

Follow the Instructions for Use of the test equipment.
When using the Testor:

Moisten the face seal with water. 

Fit the full face mask to the test head, see “Fitting the Mask” on page 3. 

Inflate test head until the sealing surface fits snugly all round.

Seal the mask connector with the adapter. 

Generate a negative pressure of 10 mbar. 

The full face mask is considered leak-proof when the pressure change after 
1 minute does not exceed 1 mbar. If not, proceed to “Testing with Sealed 
Exhalation Valve”.

Testing with Sealed Exhalation Valve

Remove the exhalation valve, and press the exhalation valve plug firmly 
onto the seat until it engages. 

Generate a negative pressure of 10 mbar. 

If the pressure change after 1 minute is less than 1 mbar, remove the 
exhalation valve plug, insert a new valve disc and repeat the leak test.
If not, proceed to “Underwater Leak Test (Optional)”.

Underwater Leak Test (Optional)

If a proper fit (leak test) cannot be achieved (no sufficient seal) with 
standard leak tests, an underwater leak test is necessary to locate the 
leak(s).

Generate an excess pressure of approx. 10 mbar. 

Immerse the test head with fitted full face mask in water and rotate slowly. 
Escaping air bubbles indicate the source of the leak(s). 

Seal leak(s). 

Remove the exhalation valve plug. 

Repeat the leak test.

After Testing

Unscrew the adapter. 

Remove the mask from the test head and dry if necessary. 

Refit the exhalation valve cover so that it clicks into place.

Order List

Approval Label

see page 14

Designation

 and description

Order No.

CDR 4500, EPDM-PC/C-RA-K/SW

R 55 440

CDR 4500, EPDM-PC/C-RA-blank

R 55 705

CDR 4500, EPDM-PC/C-RA-sw

R 55 710

CDR 4500, EPDM-PC/C-RA-S-sw

R 55 575

CDR 4500, EPDM-PC/C-RA-S-blank

R 55 715

CDR 4500, EPDM-PC/C-RA-L-sw

R 55 580

CDR 4500, EPDM-PC/C-RA-L-blank

R 55 720

Spectacle kits 

(spectacle frame and clamping device)

R 51 548

Altek spectacle kit

40 56 406

Spectacle kit PNova Trispec

40 57 622

Anti-misting gel "klar-pilot" 

(anti-misting agent for lens)

R 52 560

Blue nylon carry bag

40 55 785

Tactical mask bag

40 56 997

Mask plug

40 52 539

Airkem 33

40 54 085

Test Kit

40 58 234

Visit: www.thesafetyequipmentstore.com Or Email: [email protected] for Sales & Service.

Summary of Contents for CDR 4500

Page 1: ...500 001_CDR4500 tif NIOSH enUS Instructions for Use 2 fr Notice d utilisation 6 es Instrucciones de uso 10 Visit www thesafetyequipmentstore com Or Email besafe thesafetyequipmentstore com for Sales S...

Page 2: ...rking is externally legible and positioned on the mask DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially ha...

Page 3: ...ion Standard 29 CFR 1910 134 Failure to do so can result in serious personal injury or death 00821592_1 eps Fig 2 00221592_1 eps 1 Fig 3 00321592_1 eps 2 1 3 Fig 4 01321592_1 eps Fig 5 Positive Fit Ch...

Page 4: ...isinfection times can damage the mask 00421592_1 eps Fig 10 Maintenance Regular service maintenance and pre use checkouts of your CDR 4500 is of the utmost importance to ensure safe use of the device...

Page 5: ...for un used stored mask in original packaging otherwise every 6 months Visually inspect correct function and leak check X X1 Replace exhalation valve disc X Replace speech diaphragm X Visual check by...

Page 6: ...pas respect es il y a un possible risque de mort ou de blessure grave en raison d un danger potentiel ATTENTION Si les mesures de prudence ad quates ne sont pas respect es il y a un possible risque de...

Page 7: ...point peut entra ner des blessures graves ou mortelles 00821592_1 eps Fig 2 00221592_1 eps 1 Fig 3 00321592_1 eps 2 1 3 Fig 4 01321592_1 eps Fig 5 Contr le d ajustage positif Placez la paume de votre...

Page 8: ...92_1 eps Fig 10 Entretien Il est extr mement important d effectuer une r paration et un entretien r guliers de l quipement ainsi que de mener les contr les pr utilisation de votre CDR 4500 a n d assur...

Page 9: ...asques non utilis s et stock s dans leur emballage d origine sinon tous les 6 mois Contr le visuel contr le du bon fonctionnement et recherche de fuites X X1 Remplacement du disque de la soupape expir...

Page 10: ...correspondientes ADVERTENCIA Existe el peligro potencial de muerte o de sufrir lesiones graves si es que no se toman las medidas de precauci n correspondientes ATENCI N Existe el peligro potencial de...

Page 11: ...i n de la respiraci n En caso que no se realice esto se pueden sufrir lesiones o accidentes mortales 00821592_1 eps Fig 2 00221592_1 eps 1 Fig 3 00321592_1 eps 2 1 3 Fig 4 01321592_1 eps Fig 5 Prueba...

Page 12: ...muy alta y una exposici n prolongada a los desinfectantes puede da ar la m scara 00421592_1 eps Fig 10 Mantenimiento La realizaci n de trabajos de mantenimiento y conservaci n regular as como pruebas...

Page 13: ...co Dr ger o certi cado con el nivel II por Dr ger 00721592_1 eps Fig 13 Control visual de v lvula de exhalaci n Sacar la tapa de protecci n del conector de la m scara y revisar cuidadosamente el disco...

Page 14: ...quid chemical warfare agents dispose of the respirator after decontamination V Not for use in atmospheres immediately dangerous to life and health or where hazards have not been fully characterized W...

Page 15: ...15 Visit www thesafetyequipmentstore com Or Email besafe thesafetyequipmentstore com for Sales Service...

Page 16: ...August 2006 Subject to alteration Dr ger Safety AG Co KGaA Revalstrasse 1 D 23560 Luebeck Germany Tel 49 451 8 82 0 Fax 49 451 8 82 20 80 www draeger com Distributed By The Safety Equipment Store www...

Reviews: