![Dräger BiliLux Instructions For Use Manual Download Page 82](http://html.mh-extra.com/html/drager/bililux/bililux_instructions-for-use-manual_2526736082.webp)
82
Instructions for use BiliLux
Technical data
14.10.6
Enclosure port immunity
Test
frequency
(MHz)
Band
1)
(MHz)
1) For some services, only the up-link frequencies are included.
Service
1)
Modulation
2)
2) The carrier shall be modulated using a 50% duty cycle square wave signal.
Maximum
power
(W)
Distance
(m)
Immunity
test level
(V/m)
385
380 to 390
TETRA 400
Pulse
modulation
2)
18 Hz
1.8
0.3
27
450
430 to 470
GMRS 4460,
FRS 460
FM
3)
± 5 kHz deviation
1 kHz sine
3) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because, while it
does not represent actual modulation, it would be worst case.
2
0.3
28
710
704 to 787
LTE Band 13,
17
Pulse
modulation
2)
217 Hz
0.2
0.3
9
745
780
810
800 to 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
2)
18 Hz
2
0.3
28
870
930
1720
1700 to 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
2)
217 Hz
2
0.3
28
1845
1970
2450
2400 to 2570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
band 7
Pulse
modulation
2)
217 Hz
2
0.3
28
5240
5100 to 5800 WLAN 802.11
a/n
Pulse
modulation
2)
217 Hz
0.2
0.3
9
5500
5785
NOTE: If necessary to achieve the immunity test level, the distance between the transmitting antenna
and the ME equipment of ME system may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
Summary of Contents for BiliLux
Page 2: ...2 Instructions for use BiliLux This page has been left blank intentionally ...
Page 6: ...6 Instructions for use BiliLux This page has been left blank intentionally ...
Page 8: ...8 Instructions for use BiliLux This page has been left blank intentionally ...
Page 18: ...18 Instructions for use BiliLux This page has been left blank intentionally ...
Page 26: ...26 Instructions for use BiliLux This page has been left blank intentionally ...
Page 30: ...30 Instructions for use BiliLux This page has been left blank intentionally ...
Page 36: ...36 Instructions for use BiliLux This page has been left blank intentionally ...
Page 54: ...54 Instructions for use BiliLux This page has been left blank intentionally ...
Page 58: ...58 Instructions for use BiliLux This page has been left blank intentionally ...
Page 64: ...64 Instructions for use BiliLux This page has been left blank intentionally ...
Page 70: ...70 Instructions for use BiliLux This page has been left blank intentionally ...
Page 84: ...84 Instructions for use BiliLux This page has been left blank intentionally ...
Page 88: ...88 Instructions for use BiliLux This page has been left blank intentionally ...
Page 89: ...Instructions for use BiliLux 89 This page has been left blank intentionally ...