![Dräger BiliLux Instructions For Use Manual Download Page 68](http://html.mh-extra.com/html/drager/bililux/bililux_instructions-for-use-manual_2526736068.webp)
68
Instructions for use BiliLux
Service
12.6.3
Tightening the spring arm joints
1
Position the device so that the spring arm is in a horizontal position.
2
Remove the covers (A) on the spring arm joints.
3
Tighten the hex nut (B) in each joint using a 13 mm socket wrench so that the
device stays securely in the horizontal position. Do not overtighten.
4
Verify that the spring arm moves freely and remains in any position.
5
Reattach the covers on the spring arm joints.
12.7
Repair
Dräger recommends that all repairs are performed by DrägerService and that only
authentic Dräger repair parts are used.
03
9
A
A
B
Summary of Contents for BiliLux
Page 2: ...2 Instructions for use BiliLux This page has been left blank intentionally ...
Page 6: ...6 Instructions for use BiliLux This page has been left blank intentionally ...
Page 8: ...8 Instructions for use BiliLux This page has been left blank intentionally ...
Page 18: ...18 Instructions for use BiliLux This page has been left blank intentionally ...
Page 26: ...26 Instructions for use BiliLux This page has been left blank intentionally ...
Page 30: ...30 Instructions for use BiliLux This page has been left blank intentionally ...
Page 36: ...36 Instructions for use BiliLux This page has been left blank intentionally ...
Page 54: ...54 Instructions for use BiliLux This page has been left blank intentionally ...
Page 58: ...58 Instructions for use BiliLux This page has been left blank intentionally ...
Page 64: ...64 Instructions for use BiliLux This page has been left blank intentionally ...
Page 70: ...70 Instructions for use BiliLux This page has been left blank intentionally ...
Page 84: ...84 Instructions for use BiliLux This page has been left blank intentionally ...
Page 88: ...88 Instructions for use BiliLux This page has been left blank intentionally ...
Page 89: ...Instructions for use BiliLux 89 This page has been left blank intentionally ...