52
Instructions for use BiliLux
Operation
8.7
Ending phototherapy
During active therapy, press the
key on the control panel to turn off the
phototherapy light.
The LED indicators turn off and the phototherapy screen changes to show a paused
state:
– The "therapy on" symbol
changes to the "therapy paused" symbol
(A).
– The colon (":") in the elapsed treatment time (B) stops blinking.
To end therapy, press the
Reset
button (C).
To exit the paused state and resume therapy, press the
key on the control
panel again.
8.8
Using the MEDIBUS.X interface for data transfer
The MEDIBUS interface enables the user to export phototherapy data to an
external device via the Dräger MEDIBUS.X protocol. MEDIBUS.X is a
communication protocol for data transfer between the device and an external
medical or non-medical device (e.g., patient monitors, data management systems,
or computers) via an RS-232 communication port.
Devices connected to the RS-232 communication port must comply with the
specifications listed on page 73.
Note the following communication protocols before transmitting any data:
– MEDIBUS.X, Rules and Standards for Implementation (9052607)
– MEDIBUS.X, Profile Definition for Data Communication V1.n (9052608)
WARNING
Risk of patient injury
All data exported over the MEDIBUS.X interface are for information only.
►
Data must not be used as the sole basis for diagnostic or therapeutic decisions.
WARNING
Risk of electrical shock
Connecting devices to the MEDIBUS.X interface can lead to an increased leakage
current. If the protective grounding of one of these devices fails, the leakage current
may rise above permissible values.
►
Only connect with the approval of the respective device manufacturer.
►
Service personnel must check the leakage current.
►
If permissible values are exceeded, disconnect the devices from the
MEDIBUS.X interface.
033
008:48
008:48
Therapy duration
C
A
B
Summary of Contents for BiliLux
Page 2: ...2 Instructions for use BiliLux This page has been left blank intentionally ...
Page 6: ...6 Instructions for use BiliLux This page has been left blank intentionally ...
Page 8: ...8 Instructions for use BiliLux This page has been left blank intentionally ...
Page 18: ...18 Instructions for use BiliLux This page has been left blank intentionally ...
Page 26: ...26 Instructions for use BiliLux This page has been left blank intentionally ...
Page 30: ...30 Instructions for use BiliLux This page has been left blank intentionally ...
Page 36: ...36 Instructions for use BiliLux This page has been left blank intentionally ...
Page 54: ...54 Instructions for use BiliLux This page has been left blank intentionally ...
Page 58: ...58 Instructions for use BiliLux This page has been left blank intentionally ...
Page 64: ...64 Instructions for use BiliLux This page has been left blank intentionally ...
Page 70: ...70 Instructions for use BiliLux This page has been left blank intentionally ...
Page 84: ...84 Instructions for use BiliLux This page has been left blank intentionally ...
Page 88: ...88 Instructions for use BiliLux This page has been left blank intentionally ...
Page 89: ...Instructions for use BiliLux 89 This page has been left blank intentionally ...