7.2 Mounting at the space spring arm
Attention:
It is not possible to turn the
lamp around this joint (in horizontal
plane) because of its fixing with the
three screws.
To mount a lamp to the space spring arm proceed
as follows:
•
Loosen the fixing screw
1
of the retaining
sleeve
2
on the spring arm. Push the sleeve far
enough up for the thread to become visible.
Turn the screw
1
into the thread, so that the
sleeve cannot slip downwards.
•
Insert the stud into the spring arm in such a
way that the lamp is in the position illustrated
(see drawing –top view). Secure the lamp with
the segment
3
. Fasten the lamp with the
screws
4
.
•
Fix the retaining sleeve
2
with the screw
1
into
the initial position.
Caution:
Mount the lamp under all circumstances in
the shown position, since the bow could
touch the spring arm while handling the lamp.
Removing the mounting safeguard
The spring arm is locked in the centre position.
In order to permit free movement of the arm,
pull out the locking pin 5 on the side (approx.
1cm).
The lamp body must under all circumstances
be mounted beforehand.
Dismounting the lamp head from the Space
spring arm
Caution:
Before dismounting the lamp, push in the lock-
ing pin on the side of the arm.
•
Adjust the spring force as described in chapter
8.2.2.
•
Adjust the vertical limit as described in chapter
8.3.2.
Top view
Spring arm
105°
1 x segment
3 x screw
59500001
Edition 06