Ceiling and wall attachment
Dr. Mach
Lamps and Engineering
Central spring arm (until 2000)
•
Loosen fixing screw
7
and remove cover
8
. You
can see two slots.
•
Move the spring arm in a vertical direction, until
the hole nut is visible in one of the slots (even-
tually in the upper position of the arm).
•
Insert the key bolt (included in the scope of
supply) into the hole nut and turn accordingly.
Use only the enclosed key bolt, since a
screw driver could break.
Proceed as follows depending on the movement of
the lamp:
-
Lamp moves
down
on its own – turn bolt in the
direction of the „
+
“ mark.
-
Lamp moves
up
on its own – turn bolt in the
direction of the „
-
“ mark.
•
Fix cover
8
with the screw
7
.
Store the enclosed key bolt carefully
.
7
8
2 slots
59500001
Edition 06
11.03.2003 / Bak Page 38/48