Ceiling and wall attachment
Dr. Mach
Lamps and Engineering
Central spring arm (until 2000)
•
Remove plastic cover
6
at the upper side of the
spring arm. You can see one slot.
•
Move the spring arm into a position, in which
the hole nut is visible in the slot.
•
Turn the hole nut with the key bolt (included in
the scope of supply) in the direction of the „
+
“
mark, until the required height position has
been reached.
Store the enclosed key bolt carefully and fix
plastic cover 6.
6
-
✚
Operating instructions
For handling the lamp, please observe the operating instructions supplied by the manufacturer.
9. Maintenance
After mounting the ceiling/wall suspension and the lamps, the brakes between brackets and spring arms
and between brackets and axis (ceiling models) have to be adjusted depending on requirements.
The suspension fixtures have been designed and built, such as regular maintenance intervals are not ne-
cessary.
In order to keep the system easy running throughout its life span, we recommend that the hinges should
be greased once a year with an acid-free grease.
Caution: For the yearly maintenance the red safeguard angle G must be inserted and secured us-
ing cylinder screw H before the lamp is dismounted.
10. CE –mark
The products comply with the standard 93/42/EEC for medical products of the European
Community’s Council.