8. ZOOM/SAVE
8. ZOOM/SAVE
Window ON/OFF
Use this function to view or remove the zoom area selection bar. Press to
open the window and set the zoom area through the use of the ENCODER
switch
Using the ENCODER switch
Open the window. Rotate the ENCODER switch to expand the center of the
window as shown in the illustration below. Press (Zoom) to magnify the
desired area.
When the zoom window opens, the
selection area bar will appear. Select the
range using the ENCODER.
The selected range will cover the entire
horizontal area and the horizontal scale
factor will change.
Set up SAVE/Recall
Factory setup (setup1-setup10) Save,
(update, Recall, erase) Recall, erase.
Waveform SAVE/Recall Waveform(Setup1-setup10)
(update, Recall, erase) Save, recall, erase.
Save
Saves waveform/configuration settings
Recall
Recall waveform/configuration
settings
Erase
Deletes saved file
Clear waveform
Erase the recalled wareform and setting
in the screen
Upper Level
Sub menu
Explanation
Menu
USB flash
memory not
connected
USB flash
memory
connected
8-2. SAVE
Press longer than 2 second of the Key to enter SAVE mode.
This function downloads or saves waveform and configuration settings. The information
stored in the unit
’
s internal memory can be saved to the external USB flash memory device.
When USB flash memory is connected, select the menu item for automatically saving and
loading information onto the flash memory. If not, select the menu item for saving and
loading information into the unit
’
s internal memory. When saving internally, up to 5
waveforms and 7 configuration settings may be saved. When using USB flash memory,
waveform and configuration settings are saved simultaneously, and the saved file size is
approximately 1 kilobyte. (USB flash memory port is optional.)
54
Hand-held Digital Oscilloscope Meter
55