6. Oscilloscope Function
6. Oscilloscope Function
Upper Level
Sub menu
Explanation
Menu
Acquire
6 - 8.
ACQUIRE Press
enter Acquire mode.
Use this function to configure the signal acquisition mode. By changing the
waveform acquisition, the desired characterics among the signal characterics can
be easily achieved.
Mode
Normal, Average, Peak
Roll, Trigger
Roll, Trigger(Timebase is 50ms/div or slower)
Peak Detect
Use this function to display the minimum and maximum of the sampling
signal in 10ns units for each horizontal scale level without missing glitches.
It also allows verification of aliasing in the currently observed waveform.
Normal
Use this function to observe the signal from the ADC unchanged.
Average
Averages and displays the waveforms collected from the oscilloscope high
speed data collection terminal. The number of waveforms to average can be
adjusted between 2 and 256 using the ENCODER knob. A higher number of
waveforms to average displays a clearer waveform and decreases the
sensitivity to signal changes.
If the signal cannot be observed properly due to noise or other unexpected
signals, by selecting an appropriate number of waveforms in the average
mode, the overall flow of the signal can be verified with the noise removed
through averaging.
Rotate the ENCODER switch to select
the number of waveforms to average.
Increasing the number of waveforms
increases the clarity of the waveform and
decreases its sensitivity level.
40
Hand-held Digital Oscilloscope Meter