REV 1.0
WRM-10P S2 USER’S MANUAL
27
g.
The following screen will be displayed:
1.
YES
Press the
[1]
key (
YES
) if you wish to convert readings to standard temperature.
The following screen will be displayed:
Press the
[1]
key (
COPPER, Tk=234.5
) to select copper as the winding material.
Continue to step h
.
Press the
[2]
key (
ALUMINUM, Tk=225
) to select aluminum as the winding
material.
Continue to step h.
Press the
[3]
key (
MANUALLY ENTER Tk
) to manually enter the Tk value. The
following screen will be displayed:
You can increase the Tk value by 0.5
°
C increments by pressing the
[
]
key.
You can decrease the Tk value by 0.5
°
C increments by pressing the
[
]
key.
Press the
[ENTER]
key to confirm the temperature setting.
Continue to step h.
2.
NO
Press the
[2]
key (
NO
) if you do NOT wish to convert readings to standard
temperature.
Continue to step j.
T
K
:
230.0
°
C
↑↓
to Adjust Temp,
"ENTER" to Accept
WINDING MATERIAL
1.COPPER, T
K
=234.5
2.ALUMINUM, T
K
=225
3.MANUALLY ENTER T
K
CONVERT READINGS TO
STANDARD TEMP?
1. YES
2. NO