WRM-10P S2 USER’S MANUAL REV 1.0
18
e.
The following screen will be displayed:
Use the
[
]
and
[
]
keys to adjust the D.U.T. (Device Under Test) temperature by
0.5
°
C increments.
Press the
[ENTER]
key to confirm the D.U.T. temperature setting.
f.
The following screen will be displayed:
Use the
[
]
and
[
]
keys to adjust the reference temperature by 0.5
°
C increments.
Press the
[ENTER]
key to confirm the reference temperature setting.
NOTE
The WRM will calculate the equivalent resistance value at this new
temperature. Please see section 4.2 for further information about how this
value is calculated.
g.
The following warning screen will be displayed:
This warning is a reminder that the next sequence of test steps will run current through
the test load. Press any key to continue.
NOTE
If necessary, you can press the
[STOP]
key to cancel the test.
**********WARNING**********
DANGEROUS FLASH-OVER
WILL OCCUR IF CABLES
ARE DISCONNECTED!
***************************
REFERENCE TEMP:
75.0
°
C 167.0
°
F
↑↓
to Adjust Temp,
"ENTER" to Accept
D.U.T. TEMP:
25.0
°
C 77.0
°
F
↑↓
to Adjust Temp,
"ENTER" to Accept