IonPac NS1 Manual
Document No. 034024-06
Page 5 of 35
Table 2
NS1/NG1 Operating Parameters
Standard
Maximum
Column
Typical Back Pressure
Flow Rate
Flow Rate
psi (MPa)
mL/min
mL/min
NS1 Analytical
900 (6.21) - 1,400 (9.66)
1.0
3.0
NG1 Guard
200 (1.38) - 300 (2.07)
1.0
3.0
NS1 + NG1 columns
1,100 (7.59) - 1,700 (11.73)
1.0
3.0
NS1-5 µm analytical
1,700 (11.73) - 2,500 (17.24)
1.0
1.5
NG1 Guard
200 (1.38) - 300 (2.07)
1.0
1.5
NS1-5 µm + NG1 columns
1,900 (13.11) - 2,800 (19.31)
1.0
1.5