IonPac NS1 Manual
Document No. 034024-06
Page 25 of 35
SECTION 5 - TROUBLESHOOTING
The purpose of the Troubleshooting Guide is to help you solve operating problems that may arise while using IonPac NS1
columns. For more information on problems that originate with the Ion Chromatograph (IC) or the suppressor, refer to the
Troubleshooting Guide in the appropriate operator’s manual. If you cannot solve the problem on your own, contact the Dionex
North America Technical Call Center at 1-800-DIONEX-0 (1-800-346-6390).
Table 4
NS1/NG1 Troubleshooting Summary
Observation Cause
Action Reference
Section
High Back Pressure
Unknown Isolate
Blocked
Component
5.1.1
Plugged Column Bed Supports
Replace Bed Supports
5.1.2
Other System Modules
Disconnect, Replace
System Module Manual
High Background
Conductivity
Bad Eluents
Remake Eluents
5.2, 5.2.1
Contaminated
Columns
Clean
Column
5.2.2,
5.2.3
Contaminated ASRS or AMMS
Clean Suppressor
5.2.5
Contaminated Hardware
Clean Component
Poor Resolution
Poor Efficiency Due to Large
System Void Volumes
Replumb System
5.3.1.A, 5.3.3.D
Column Headspace
Replace Column
5.3.1.B
Short Retention Times
Unequilibrated System
Lengthen First Eluent
Time before Inject
5.3.3.C
Flow Rate Too fast
Recalibrate Pump
5.3.2.A
Bad
Eluents
Remake
Eluents
5.3.2.B
Column Contamination
Clean Column
5.3.2.C, 5.3.2.D
Poor Front End Resolution
Bad Eluents
Remake Eluents
5.3.3.A
Column Overloading
Reduce Sample Size
5.3.3.B, 2.2
Sluggish
Injection
Valve
Service
Valve
5.3.3.C
Large System Void Volumes
Replumb System
5.3.3.D
Spurious Peaks
Sample Contamination
Pretreat Samples
5.4.A, 5.4.B
Sluggish
Injection
Valve
Service
Valve
5.4.C