3 • Operation and Maintenance
Doc. 065215-02 9/09
29
3.4
Preparing the Eluent
1. Prepare the eluent. For instructions refer to the column manual. Column
manuals are included on the Dionex Reference Library CD-ROM (P/N
053891).
2. Fill the eluent bottle with the prepared eluent, insert the stopper assembly
tubing into the bottle, and hand tighten the cap.
Set the Eluent Level
After filling the bottle, move the
Eluent Remaining
slider on the ICS-900
) to indicate the volume of liquid in the bottle.
The ICS-900 determines the eluent usage by monitoring the flow rate and the
length of time the pump is on. As the eluent is used up, the ICS-900 updates the
Eluent Remaining
slider and gauge. A warning appears in the Audit Trail when
the level falls below 200 mL, and then again when the level falls bellow 100 mL.
Figure 3-2. ICS-900 Control Panel on Panel Tabset