ICS-900 Operator’s Manual
94
Doc. 065215-02 9/09
A.4 Front and Rear Panel LEDs
Front Panel LEDs
Rear Panel LED
A.5 Pump
Power
Indicates when the power is on.
Ready
Indicates when the ICS-900 is ready to acquire data; flashes if the
system check fails.
Run
Indicates when the ICS-900 is running/acquiring data; flashes if an
error occurs.
Link
Indicates when there is communication between the ICS-900 and
Chromeleon or Chromeleon Xpress; flashes when data is being
transmitted or received.
Type
Serial dual-reciprocating piston, microprocessor-controlled
constant stroke, variable speed
Construction
Chemically inert, metal-free PEEK pump heads and flow path.
Compatible with aqueous eluents of pH 0 to 14 and reversed-phase
solvents
Operating
Pressure
0 to 35 MPa (0 to 5000 psi)
Flow Rate
Range
0.00 to 5.00 mL/min, in 0.01 mL/min increments (without
changing pump heads)
Flow Precision
<0.1%
Flow Accuracy
<0.1%
Pressure Ripple
<1%
Eluent Bottle
Pressure
None required