D.10 How do I shut off the system?
In Chromeleon or Chromeleon Xpress, turn off the pump from the ICS-900
Control panel (see
). On the ICS-900, turn off the power switch on the
D.11 How do I store columns?
Columns should be stored in eluent. See the column manual for complete
instructions. Column manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891).
D.12 How do I know when a column is dirty?
See the troubleshooting section of the column manual. Column manuals are
included on the Dionex Reference Library CD-ROM (P/N 053891).
D.13 How do I clean a column?
See the troubleshooting section of the column manual. Column manuals are
included on the Dionex Reference Library CD-ROM (P/N 053891).
D.14 Why is the conductivity high?
Possible reasons for high conductivity include:
•
The suppressor needs regeneration. See the suppressor manual for
troubleshooting information. Suppressor manuals are included on the Dionex
Reference Library CD-ROM (P/N 053891).
•
The regenerant is exhausted and should be remade. See
for
instructions on preparing regenerant.
•
The cell is out of calibration. See
for calibration instructions.