2. ERROR MESSAGES
The errors that may appear during the different processes that the analyzer runs are:
Error
Description
Possible Cause
Effect
Action
Non-
CALIB
Uncalibrated
electrode. The
name of the
Electrode
is
displayed
in
Red
Check the profit
and
symbols
behind
the
phrase
"No
CALIB"
The parameter
will
not
be
measured.
● Check if there are
other symbols in
the calibration.
● Retry Calibrate.
● Change the Pack.
● Change
the
electrode.
/
The
measurement
made is not
stable
Bubbles in the
liquid circuit.
Interference
during
stabilization.
Electrode failure.
Failure
in
the
measurement
channel.
Unstable sample.
In calibration:
The electrolyte
will
not
be
measured (not
calib.).
In
measurement:
The result is
reported
but
will
not
be
valid.
● Repeat calibration /
measurement.
● Check plugs on
pipes
and
electrodes.
● Eliminate
any
possible sources of
interference.
● Check the valves
and peristaltic pipe.
In
the
case of
electrodes
with
glass capillary (Na /
pH), pass sodium
conditioner.
● Possible electronic
failure.
● Change electrode.
● If
the
problem
persists,
call
technical support.
?
ADC
out
of
range.
Bubbles in the
liquid circuit.
Interference
during
stabilization.
Electrode failure.
Failure
in
the
measurement
channel.
In calibration:
The electrolyte
will
not
be
measured.
In
the
measurement:
The result will
not be valid.
S
Low or High
Gain.
Bubbles in the
liquid circuit.
Interference
during
stabilization.
Electrode failure.
Failure
in
the
measurement
channel.
The electrolyte
will
not
be
measured.
The calibration
slope
is
Calibration
solutions
are
Electrolytes
cannot
be
● Verify the correct
connection of the
Summary of Contents for 103APV4R
Page 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Page 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Page 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Page 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...