DIESTRO 103APV4R ELECTROLYTE ANALYZER
20
JS Medicina Electrónica
2. MICROCHIP (uChip)
Pack
’s uChip
The uChip integrated in the Pack/Kit provides the analyzer with information about the volumes
of solutions, calibration values, content, type, batch and expiration date of the Pack.
3. VALIDITY OF THE PACK
Check the expiration date of the pack to be connected
.
In the event that the pack is expired, the analyzer will display and print the message Pack
Expired and the
status light flashes green to alert the user. Even so, you can use the equipment, under the sole
responsibility of the operator.
4. CONSUMPTION OF CALIBRATOR SOLUTIONS
The analyzer electronically discounts "doses" from the uChip.
When the pack is empty, the analyzer will display the message "Pack Empty", and will stop
working with that pack.
When the pack is empty, you will need to discard it and install a new one.
The red cap and associated tubing of the Analyzer may contain potentially
infectious residues, use caution, wear gloves, do not splash.
The depleted pack contains potentially infectious residues. Discard the pack
following the clarifications in "WASTE DISPOSAL"
Take care not to contaminate the green and orange lids and their respective tubing,
as they are used for the next pack
.
To avoid spills, place the green, orange and red caps into their respective
containers before discarding.
Summary of Contents for 103APV4R
Page 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Page 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Page 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Page 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...