DIESTRO 103APV4R ELECTROLYTE ANALYZER
107 JS Medicina Electrónica
Reconnect the analyzer outlet and check in the sampler settings menu that the source position
is correct. If it is not correct, correct it in the Autosampler settings menu.
10. CHANGE SAMPLER DISC SEPARATORS
In the case of using tubes of 75mm long or shorter, it is recommended to change the separators
of the sampler disc for the shorter ones provided with the analyzer. To make this change
perform the following sequence.
- Remove the sampler disc as indicated in the previous section.
- Remove the bottom disc with the supplied Allen M2.5 wrench.
- Remove and replace the long separators with the shorter ones.
- Screw the bottom disc back in.
- Replace the sampler disc.