ION Calibration Result Visualization
TCO2 calibration result display (if available)
You can print these results by pressing the Print button.
The legend "Calibrated" indicates that the electrode is ready to measure.
See the "ELECTRODE GAIN" section for gain ranges
If the sodium or pH gain is less or greater than expected, perform a wash with
sodium conditioner. For more details see the section "SODIUM / PH
CONDITIONER".
The legend "Not Calibrated" indicates that the electrode will not be able to
measure. View the "ERROR MESSAGES" section