1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 43 of 58
12. Connect the waste tube from the TPR block to the Peristaltic Pump.
13. Connect the reference out tube to the Peristaltic Pump.
14. Connect the reference tube from the reference solution connector to the peristaltic pump.
Ensure the reference tubing’s are connected correctly. Misplacement will cause malfunction
of the reference electrode.
15. Slide the TPR block into place.
16. Pull tubes out below the TPR block on left side.
17. Install the reference tube and reference solution tube connector.
18. Place tubing into Left Guide Track, with shortest first (Waste In) to the longest, Standard A.
19. Place all Standard A, B, C, Reference solution and Air tubings into their tracks followed by
their respective crimp pin then the indicator cap.
20. Snap peristaltic pump into place.
21. Connect the red tagged tube to the bottom of the fill port.
22. Connect the green tagged tube to the Electrode Tray.
23. Install the Fluid Pack and run Prime All to fill reagent lines.
(Main Operations Menu>>>Maintenance>>>Prime All).
Figure 1-61 Pinch Valve Tubing Set Diagram
TPR
Green
Tube
Red
Tube
Peristaltic
Pump Tubing
Blue
Tube
Tubing
Cradle