1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 13 of 58
Figure 1-6 Door Hinge Attachment Details
4. With one hand, hold the analyzer door near the right side hinge pin. (
5. Ensure the door is not on the work surface.
6. With the other hand, gently apply pressure to the middle rear area of the door. This will flex
the door slightly and allows the right hinge pin to clear the retaining hole in the main chassis.
7. Remove the right hinge pin then the left so the door can then be removed from the analyzer.
Figure 1-7 Door Details for Removal
The window is comprised of two layers, the EMI window and the plastic window. Both can be
replaced in the field.
1. From the rear side of the door, (
), gently press the window and snapping the EMI
and plastic window out the front side of the door.
2. Verify that the copper conductive tape is still adhered to the both the front and back surface
of the door.
3. Remove adhesive backing from the EMI Window.
4. Place the EMI Window first over the opening on the front side of the door.
5. Press along the frame of the window and ensure good contact of the EMI window with the
copper tape on the frame of the window.
6. After removing the adhesive backing, apply the plastic window over the EMI window pressing
along the frame again to ensure good contact.
Ground Screw
External Tooth
Washer
Right Angle Tab
Hex Nut
Apply Gentle
Pressure
Right Hinge