3. OPERATION
Page 77
© 2007 DH Instruments, a Fluke Company
3.5.4.3
<3AT
EST
>
PURPOSE
•
Resets Quick Test and Q-DUT Tests to default (see Section 3.3.6).
•
Clears AutoTest data log (see Section 3.4.5).
3.5.4.4
<4CAL>
PURPOSE
The Reset - Cal function will reset Q-RPT, utility sensor and barometer
calibration coefficients and settings and reset AutoZ values to zero. This will change
the PPCH-G calibration and could cause it to make out of tolerance measurements.
Clears all user values affecting the calibration of Q-RPTs, utility sensor and the
on-board barometer. Does not clear factory coefficients. The Reset – Cal resets
are itemized in Table 11.
Table 11.
Reset – Cal
RESET
RESULT
SEE SECTION
All Q-RPT user calibration
coefficients
PA to zero, PM to 1.
5.2.1.1, 5.2.7
Q-RPT absolute mode
ON.
5.2.5
Utility Sensor calibration
coefficients
PA to zero, PM to 1.
5.4
On-board barometer
calibration coefficients
PA to zero, PM to 1.
5.3
Calibration date
Set all dates to 19800101.
5.2.7, 5.4, 5.3
AutoZ values
All P
offset
values to zero for absolute mode and
101325 Pa for gauge modes.
3.5.1
AutoZ function
ON, all Q-RPTs, all measurement modes.
3.5.1
3.5.4.5
<5ALL>
PURPOSE
The reset - all function clears and deletes large amounts of user defined
information including critical calibration data.
Combines all resets in one global reset command that clears the entire user
section of non-volatile memory returning PPCH-G to the “as delivered” condition
except the ID function (see Section
3.5.5.4) and the security level password (see
Section
3.5.5.5). The Reset – All resets are itemized in Table 12.